Instrument Functions
R&S
®
FSVA/FSV
395Operating Manual 1176.7510.02 ─ 09
SWE:EGAT:TRAC1:STAT1 ON (see [SENSe:]SWEep:EGATe:TRACe<k>[:
STATe<range>] on page 855)
Activates tracing for range 1 of trace 1.
SWE:EGAT:TRAC1:STAR1 3ms (see [SENSe:]SWEep:EGATe:TRACe<k>:
STARt<range> on page 856)
Sets the starting point for range 1 on trace 1 at 3 ms.
SWE:EGAT:TRAC1:STop1 5ms (see [SENSe:]SWEep:EGATe:TRACe<k>:
STOP<range> on page 856)
Sets the stopping point for range 1 on trace 1 at 5 ms.
SWE:EGAT:TRAC1:PER 5ms (see [SENSe:]SWEep:EGATe:TRACe<k>: PERiod
on page 856)
Defines the period for gated triggering to 5 ms.
Adjust Settings ← APD
Adjusts the level settings according to the measured difference between peak and min-
imum power for APD measurement or peak and mean power for CCDF measurement
in order to obtain maximum power resolution. Adjusts the reference level to the current
input signal. For details see also the Adjust Ref Lvl softkey.
Remote command:
CALCulate<n>:STATistics:SCALe:AUTO ONCE on page 739
Softkeys for CCDF Measurements
CCDF.......................................................................................................................... 395
Percent Marker............................................................................................. 396
Res BW.........................................................................................................396
# of Samples.................................................................................................396
Scaling.......................................................................................................... 397
x-Axis Ref Level..................................................................................397
x-Axis Range...................................................................................... 397
Range Log 100 dB....................................................................397
Range Log 50 dB......................................................................397
Range Log 10 dB......................................................................397
Range Log 5 dB........................................................................398
Range Log 1 dB........................................................................398
Range Log Manual................................................................... 398
Range Linear %........................................................................398
Range Lin. Unit.........................................................................399
y-Axis Max Value................................................................................399
y-Axis Min Value.................................................................................399
y-Unit % / Abs.....................................................................................399
Default Settings.................................................................................. 399
Adjust Settings....................................................................................400
Gated Trigger (On/Off)..................................................................................400
Gate Ranges.................................................................................................400
Adjust Settings..............................................................................................401
CCDF
Activates the function to measure the complementary cumulative distribution function
(CCDF) and opens a submenu.