[SENSe:]MIXer:SIGNal <State>
This command specifies whether automatic signal detection is active or not.
Parameters:
<State> OFF | ON | AUTO | ALL
OFF
No automatic signal detection is active.
ON
Automatic signal detection (Signal ID) is active.
AUTO
Automatic signal detection (Auto ID) is active.
ALL
Both automatic signal detection functions (Signal ID+Auto ID)
are active.
*RST: OFF
Example:
MIX:SIGN AUTO
Mode: all
Manual operation: See "Signal ID (On/Off)" on page 533
See "Auto ID (On/Off)" on page 534
[SENSe:]MIXer:THReshold <Value>
This command defines the maximum permissible level difference between test sweep
and reference sweep to be corrected during automatic comparison (see [SENSe:
]MIXer:SIGNal on page 829).
Parameters:
<Value> <numeric value>
Range: 30 dBm to -130 dBm
*RST: 0.0 dBm
Example:
MIX:PORT 3
Mode: all
Manual operation: See "Auto ID Threshold" on page 535
SENSe:MPOWer Subsystem
The commands of this subsystem are used to determine the mean burst power or peak
burst power for a given number of signal bursts, and for outputting the results in a list.
Since all the settings required for a measurement are combined in a single command,
the measurement speed is considerably higher than when using individual commands.
For measuring the signal bursts, the gated sweep function is used in zero span. The
gate is controlled either by an external trigger signal or by the video signal. An individ-
ual trigger event is required for each burst to be measured. If an external trigger signal