Remote Control
R&S
®
FSVA/FSV
814Operating Manual 1176.7510.02 ─ 09
<AVGPower> ON | OFF
Turns average power evaluation on and off.
*RST: OFF
<TriggerSource> IMMediate | EXTernal | VIDeo | IFPower | RFPower
Selects a trigger source.
For more information see "Trg/Gate Source" on page 268.
*RST: IMM
<TriggerSlope> NEGative | POSitive
Selects the trigger slope.
*RST: POS
<TriggerOffset> Defines the trigger delay. The trigger delay is the time between
the detection of the trigger signal and the start of the measure-
ment at the next frequency point.
Range: span = 0 Hz: (-sweep time) to 30 s; span > 10 Hz:
31.25 ns to 30 s
*RST: 0 s
<GateLength> Defines the gate length for gated measurements.
Setting 0 seconds turns gated measurements off. Other values
activate the gated trigger function.
To perform gated measurements, the trigger source must be dif-
ferent from IMMediate.
Range: 31.25 ns to 30 s
*RST: 0 s
Example:
SENSe:LIST:POWer:SET ON,OFF,OFF,EXT,POS,10US,
434US
[SENSe:]LIST:POWer:STATe <State>
This command turns the List Evaluation off.
Parameters:
<State> OFF
Example:
SENS:LIST:POW:STAT OFF
SENSe:LIST:RANGe Subsystem
The SENSe:LIST:RANGe Subsystem contains the remote commands to configure
Spurious Emissions measurements.
[SENSe:]LIST:RANGe<range>:BANDwidth[:RESolution]....................................................815
[SENSe:]LIST:RANGe<range>:BANDwidth:VIDeo.............................................................815
[SENSe:]LIST:RANGe<range>:BREak.............................................................................815
[SENSe:]LIST:RANGe<range>:COUNt.............................................................................816
[SENSe:]LIST:RANGe<range>:DELete............................................................................ 816
[SENSe:]LIST:RANGe<range>:DETector......................................................................... 816