3.9.5 Power Sensor Configuration Dialog
Power sensors are configured in the "Power Sensor Configuration" dialog box, which is
displayed when you select the "Power Sensor Config" softkey in the "Power Sensor"
menu.
Up to 4 different power sensors can be configured. Each sensor configuration is dis-
played on a separate tab. The serial number of the power sensor connected to the
instrument and currently assigned to the displayed configuration is indicated on the
tab.
For details on how to configure and activate power sensors, see Chapter 3.9.1, "Using
Power Sensors", on page 541.
Select
If enabled, the power sensor is evaluated when power measurement is activated.
Remote command:
[SENSe:]PMETer<p>[:STATe] on page 834
Zeroing Power Sensor
Starts zeroing of the power sensor. For details on the zeroing process refer to Chap-
ter 3.9.2, "Zeroing the Power Sensor", on page 542 .
Remote command:
CALibration:PMETer<p>:ZERO:AUTO ONCE on page 744
Frequency Manual
The frequency of the signal to be measured. The power sensor has a memory with fre-
quency-dependent correction factors. This allows extreme accuracy for signals of a
known frequency.