EasyManua.ls Logo

Rohde & Schwarz FSVA - 4.2.4.10 SOURce Subsystem; SOURce:TEMPerature Subsystem

Rohde & Schwarz FSVA
1106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSVA/FSV
955Operating Manual 1176.7510.02 ─ 09
[SENSe:]ROSCillator:SOURce <Source>
This command selects the reference oscillator.
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Parameters:
<Source> INTernal
The internal reference is used (10 MHz)
EXTernal
The external reference is used; if none is available, an error flag
is displayed in the status bar
EAUTo
The external reference is used as long as it is available, then the
instrument switches to the internal reference
Example:
ROSC:SOUR EXT
Usage: SCPI confirmed
Manual operation: See "Reference Int/Ext" on page 173
See "Show Error Flag" on page 174
See "Auto select Reference" on page 174
[SENSe:]ROSCillator:SOURce:EAUTo?
This command queries the current reference type in case you have activated an auto-
matic switch to the internal reference if the external reference is missing.
Return values:
<Reference> INT
internal reference
EXT
external reference
Example:
SENS:ROSC:SOUR:EAUT?
Queries the currently available reference type.
Usage: Query only
4.2.4.10 SOURce Subsystem
The SOURce subsystem controls the output signals of the instrument.
SOURce:TEMPerature Subsystem
SOURce<n>:TEMPerature:APRobe
This command queries the sensor temperature.
Suffix:
<n>
.
irrelevant
Remote Control – Commands

Table of Contents