Remote Control
R&S
®
FSVA/FSV
659Operating Manual 1176.7510.02 ─ 09
CALCulate<n>:LIMit<k>:LOWer:STATe <State>
This command turns a lower limit line on and off.
Before you can use this command, you have to select a limit line with
CALCulate<n>:LIMit<k>:NAME.
The limit check is activated separately with CALCulate<n>:LIMit<k>:STATe. The
result of the limit check can be queried with CALCulate<n>:LIMit<k>:FAIL?.
Suffix:
<n>
.
irrelevant
<k> 1...8
Selects the limit line.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:LIM4:LOW:STAT ON
Switches on limit line 4 (lower limit).
CALCulate<n>:LIMit<k>:LOWer:THReshold <Threshold>
This command defines a threshold for relative limit lines.
The R&S FSVA/FSV uses the threshold for the limit check, if the limit line violates the
threshold.
Suffix:
<n>
.
irrelevant
<k> 1...8
Selects the limit line.
Parameters:
<Threshold> Numeric value that defines the threshold.
The unit depends on CALCulate<n>:LIMit<k>:UNIT.
*RST: -200 dBm
Example:
CALC:LIM2:LOW:THR -35DBM
Defines an absolute threshold value for limit line 2.
CALCulate:LIMit:UPPer Subsystem
The CALCulate:LIMit:UPPer subsystem defines the upper limit line.
CALCulate<n>:LIMit<k>:UPPer[:DATA]............................................................................660
CALCulate<n>:LIMit<k>:UPPer:MARGin.......................................................................... 660
CALCulate<n>:LIMit<k>:UPPer:MODE.............................................................................661
CALCulate<n>:LIMit<k>:UPPer:OFFSet...........................................................................661
CALCulate<n>:LIMit<k>:UPPer:SHIFt.............................................................................. 661
Remote Control – Commands