Remote Control
R&S
®
FSVA/FSV
766Operating Manual 1176.7510.02 ─ 09
Parameters:
<State> ON
User port is used as a input.
OFF
User port is used as a output.
*RST: ON
Example:
INP:UPOR:STAT ON
4.2.3.10 INSTrument Subsystem
The INSTrument subsystem selects the operating mode of the unit either via text
parameters or fixed numbers.
INSTrument[:SELect] <Mode> | <ChannelName>
This command selects the measurement mode by means of text parameters.
In case you are working with more than one spectrum display, the command also
selects the spectrum display (or channel) you need.
Note: To select an I/Q Analyzer channel in simple mode, you have to use TRACe<n>:
IQ[:STATe].
Also see
INSTrument:CREate[:NEW] on page 767
Remote program example: Chapter 4.3.16, "Usage of Four Spectrum Instances",
on page 1049
Parameters:
<Mode> values see Table 4-14
*RST: SANalyzer
Example:
INST SAN
Switches the instrument to "Spectrum" mode.
Usage: SCPI confirmed
Table 4-14: Instrument mode parameters
Mode Text parameter
(INST:SEL)
Numeric
parameter
(INST:NSEL)
Option number
Spectrum SANalyzer 1
Additional spectrum mode.
The first spectrum mode is
always active (SANalyzer).
Spectrum 2 |
Spectrum 3 |
Spectrum 4
IQ Analyzer (eval mode) IQ -
Analog demodulation ADEMod 3 R&S FSV-K7
FM Stereo SFM 7 R&S FSV–K7S
Bluetooth BTOoth 12 R&S FSV-K8
Remote Control – Commands