Instrument Functions
R&S
®
FSVA/FSV
451Operating Manual 1176.7510.02 ─ 09
The measurement is performed on the trace that marker 1 is assigned to. To shift
marker 1 and measure another trace, use the Marker to Trace softkey in the "Marker
To" menu.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:POWer:SELect on page 703
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult? on page 700
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe] on page 704
Channel Bandwidth ← C/N, C/No ← I/Q Analyzer
Opens an edit dialog box to enter the measurement channel bandwidth for each chan-
nel.
The default setting is 14 kHz.
Note that in the I/Q Analyzer the channel bandwidth does not change the measure-
ment span or sampling rate; it merely defines the range for the carrier-to-noise analy-
sis.
Remote command:
[SENSe:]POWer:ACHannel:ACPairs on page 842
Adjust Settings ← C/N, C/No ← I/Q Analyzer
Enables the RMS detector (see also Chapter 3.2.8.6, "Detector Overview",
on page 265) and adjusts the span to the selected channel bandwidth according to:
"4 x channel bandwidth + measurement margin"
The adjustment is performed once; if necessary, the setting can be changed later on.
Note: in the I/Q Analyzer, this function is not available. The channel bandwidth defines
the range for the carrier-to-noise analysis.
Remote command:
[SENSe:]POWer:ACHannel:PRESet on page 847
3.5.2 Softkeys of the Amplitude Menu in I/Q Analyzer Mode
In I/Q Analyzer mode, the "Amplitude" menu, which is displayed when you select the
AMPT key, contains the following functions.
If the display configuration for the I/Q Analyzer is set to "I/Q Vector" or "Real/Imag (I/
Q)", the Range and Unit functions are not available.
Ref Level.....................................................................................................................452
Y-Axis Max..................................................................................................................452
Range..........................................................................................................................452
Range Log 100 dB........................................................................................452
Range Log 50 dB..........................................................................................452
Range Log 10 dB..........................................................................................453
Range Log 5 dB............................................................................................453
Range Log 1 dB............................................................................................453
Range Log Manual........................................................................................453
Instrument Functions - I/Q Analyzer