Parameters:
<MeasType> ACPower | CPOWer | MCACpower | OBANdwidth | OBWidth |
CN | CNO
ACPower
Adjacent-channel power measurement with a single carrier sig-
nal
CPOWer
Channel power measurement with a single carrier signal (equiv-
alent to adjacent-channel power measurement with "NO. OF
ADJ CHAN" = 0)
MCACpower
Channel/adjacent-channel power measurement with several car-
rier signals
OBANdwidth | OBWidth
Measurement of occupied bandwidth
CN
Measurement of carrier-to-noise ratio
CN0
Measurement of carrier-to-noise ratio referenced to 1 Hz band-
width
Example:
CALC:MARK:FUNC:POW:SEL ACP
Switches on adjacent-channel power measurement.
Manual operation: See "Ch Power ACLR" on page 311
See "OBW (span > 0)" on page 312
See "Ch Power ACLR" on page 324
See "C/N" on page 339
See "C/No" on page 339
See "OBW" on page 343
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe] <State>
This command switches off the power measurement in the window specified by the
suffix <n>.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<State> OFF
Example:
CALC:MARK:FUNC:POW OFF
Switches off the power measurement.
Usage: Event