Quick Start
R&S
®
FSVA/FSV
17Operating Manual 1176.7510.02 ─ 09
The female connector and volume control are available only with the audio demodula-
tor option (R&S FSV-B3). To use the AF OUTPUT, in the "In-/Output" menu (INPUT/
OUTPUT key), select "Video Output". The output voltage (volume) is 1V.
Risk of hearing damage
To protect your hearing, make sure that the volume setting is not too high before put-
ting on the headphones.
Tracking Generator Output (GEN OUTPUT 50Ω, Option R&S FSV-B9)
The R&S FSVA/FSV is equipped with an optional tracking generator and therefore pro-
vides a tracking generator output connector.
The output of the tracking generator is to be connected to the DUT via a cable equip-
ped with a male N connector. The female connector is available only with the tracking
generator option (R&S FSV-B9).
If the output is active, the LED is green.
Risk of damaging the tracking generator
To avoid damaging the tracking generator, make sure that the reverse power (i.e. input
power caused, for example, by reflections in the test setup) never exceeds 1 W RF
power. Do not apply a DC voltage at the output.
For details, refer to the data sheet.
Sensitive DUTs concerning matching
For DUTs with sensitive RF characteristics with regard to matching (VSWR) at the
input, insert a 10 dB attenuator between the DUT and the tracking generator.
Connections for External Mixers (EXT MIXER, Option R&S FSV-B21)
External mixers can be connected at the LO OUT/IF IN and IF IN female connectors
(option R&S FSV-B21 for R&S FSVA/FSV 30 and R&S FSVA/FSV 40 instruments).
Both two-port and three-port mixers can be used. Connect the mixer as follows:
Use the supplied coaxial cable to feed in the LO signal. If no external mixers are con-
nected to the R&S FSVA/FSV, cover the two front connectors LO OUT / IF IN and IF
IN with the SMA caps supplied.
Front and Rear Panel View