Quick Start
R&S
®
FSVA/FSV
18Operating Manual 1176.7510.02 ─ 09
Three-port mixer
1. Connect the LO OUT / IF IN output of the R&S FSVA/FSV to the LO port of the
external mixer.
2. Connect the IF IN input of the R&S FSVA/FSV to the IF port of the external mixer.
3. Feed the signal to be measured to the RF input of the external mixer.
Two-port mixer
1. 1. Connect the LO OUT / IF IN output of the R&S FSVA/FSV to the LO/IF port of
the external mixer. The nominal LO level is 15.5 dBm.
Because of the diplexer contained in the R&S FSVA/FSV, the IF signal can be tap-
ped from the line which is used to feed the LO signal to the mixer.
Front and Rear Panel View