Instrument Functions
R&S
®
FSVA/FSV
195Operating Manual 1176.7510.02 ─ 09
2. Press the "Rescan" button.
3. Press the "Save" button.
The instrument reboots and after the reboot LXI is active.
4. To deactivate the LXI Class C functionality perform step 1 and 2 again.
An LXI configuration dialog box is displayed.
5. Press the "Turn LXI Off" button.
3.1.1.3 LXI Class C Functionality
If the LXI Class C functionality is installed and enabled (default state is off; see Chap-
ter 3.1.1.1, "Softkeys of the Setup Menu", on page 171), the instrument can be
accessed via any web browser (e.g. the Microsoft Internet Explorer) to perform the fol-
lowing tasks:
●
modifying network configurations
●
modifying device configurations
●
monitoring connections from the device to other devices
To change settings, in the web browser, open the "http://<instrument-hostname>" or
"http://<instrument-ip-address>" page. The password to change LAN configurations is
LxiWeb.
General Settings, Printout and Instrument Settings