EasyManua.ls Logo

Rohde & Schwarz FSVA - CALCulate:FLINe Subsystem

Rohde & Schwarz FSVA
1106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSVA/FSV
635Operating Manual 1176.7510.02 ─ 09
Suffix:
<n>
.
Selects the measurement window.
Parameters:
<State> ON | OFF
*RST: ON
Example:
CALC:ESP:PSE:AUTO OFF
Deactivates the list evaluation.
CALCulate<n>:ESPectrum:PSEarch|:PEAKsearch:MARGin <Margin>
This command sets the margin used for the limit check/peak search.
Suffix:
<n>
.
Selects the measurement window.
Parameters:
<Margin> -200 to 200 dB
*RST: 200 dB
Example:
CALC:ESP:PSE:MARG 100
Sets the margin to 100 dB.
CALCulate<n>:ESPectrum:PSEarch|:PEAKsearch:PSHow <State>
This command marks all peaks with blue squares in the diagram.
Suffix:
<n>
.
Selects the measurement window.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:ESP:PSE:PSH ON
Marks all peaks with blue squares.
Manual operation: See "Show Peaks" on page 352
CALCulate:FLINe Subsystem
The CALCulate:FLINe subsystem defines the position of the frequency lines.
CALCulate<n>:FLINe<k>................................................................................................635
CALCulate<n>:FLINe<k>:STATe.....................................................................................636
CALCulate<n>:FLINe<k> <Frequency>
This command defines the position of a frequency line.
Suffix:
<n>
.
Selects the measurement window.
Remote Control – Commands

Table of Contents