Remote Control
R&S
®
FSVA/FSV
635Operating Manual 1176.7510.02 ─ 09
Suffix:
<n>
.
Selects the measurement window.
Parameters:
<State> ON | OFF
*RST: ON
Example:
CALC:ESP:PSE:AUTO OFF
Deactivates the list evaluation.
CALCulate<n>:ESPectrum:PSEarch|:PEAKsearch:MARGin <Margin>
This command sets the margin used for the limit check/peak search.
Suffix:
<n>
.
Selects the measurement window.
Parameters:
<Margin> -200 to 200 dB
*RST: 200 dB
Example:
CALC:ESP:PSE:MARG 100
Sets the margin to 100 dB.
CALCulate<n>:ESPectrum:PSEarch|:PEAKsearch:PSHow <State>
This command marks all peaks with blue squares in the diagram.
Suffix:
<n>
.
Selects the measurement window.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:ESP:PSE:PSH ON
Marks all peaks with blue squares.
Manual operation: See "Show Peaks" on page 352
CALCulate:FLINe Subsystem
The CALCulate:FLINe subsystem defines the position of the frequency lines.
CALCulate<n>:FLINe<k>................................................................................................635
CALCulate<n>:FLINe<k>:STATe.....................................................................................636
CALCulate<n>:FLINe<k> <Frequency>
This command defines the position of a frequency line.
Suffix:
<n>
.
Selects the measurement window.
Remote Control – Commands