Quick Start
R&S
®
FSVA/FSV
37Operating Manual 1176.7510.02 ─ 09
If you do not understand the currently used interface language, this may be difficult.
Thus, the position of the required elements in the menu is provided in the following pro-
cedure for convenience.
1. Select the SETUP hardkey.
2. Select the "General Settings" softkey (5th from the top).
3. Select the "Language" softkey (6th from the top).
4. Select an available interface language.
All softkeys, dialog boxes, input settings etc. are displayed in the selected lan-
guage.
To change the default input language of an external keyboard, use the standard Win-
dows operating system function (see "Connecting a keyboard" on page 35).
Remote command:
SYSTem:DISPlay:LANGuage on page 965
2.2.4.2 Selecting the Frequency Reference Signal
You can switch the reference signal for frequency processing of the R&S FSVA/FSV
between the internal reference and an external reference signal at 10 MHz as follows:
1. Press the SETUP key.
2. Press the "Reference Int/Ext" softkey until it is in the desired state.
External reference signal
It is important that the external reference signal is deactivated when switching from
external to internal reference to avoid interactions with the internal reference signal.
Remote commands:
ROSC:SOUR EXT
ROSC:EXT:FREQ 20
2.2.4.3 Setting the Date and Time
You can set the date and time for the internal real time clock as follows:
Opening the Date and Time Properties dialog box
1. Press the SETUP key.
2. Press the "General Setup" softkey.
Preparing for Use