To start a new channel
1. Click the icon in the toolbar at the top of the screen (see also Chapter 2.4.2.1,
"Toolbar", on page 69 on how to display the toolbar).
2. Select the softkey for the required measurement mode.
A new tab is displayed for the new channel.
2.4.1.2 Display of Hardware Settings
Information on hardware settings are displayed in the channel bar above the diagram.
Invalid settings
A bullet next to the hardware setting indicates that user-defined settings are used, not
automatic settings. A green bullet indicates this setting is valid and the measurement is
correct. A red bullet indicates an invalid setting that does not provide useful results.
It is the user's responsibility to remedy such a situation.
Editing settings in the channel bar
All settings that are displayed in the channel bar can easily be edited by touching the
setting in the display (with a finger or mouse pointer). The corresponding (edit) dialog
box is displayed where you can edit the setting.
In spectrum mode, the R&S FSVA/FSV shows the following settings:
Ref Level Reference level
m.+el.Att Mechanical and electronic RF attenuation that has been set.
Ref Offset Reference level offset
SWT Sweep time that has been set.
If the sweep time does not correspond to the value for automatic coupling,
a bullet is displayed in front of the field. The color of the bullet turns red if
the sweep time is set below the value for automatic coupling. In addition,
the UNCAL flag is shown. In this case, the sweep time must be increased.
RBW Resolution bandwidth that has been set.
If the bandwidth does not correspond to the value for automatic coupling,
a green bullet appears in front of the field.
VBW Video bandwidth that has been set.
If the bandwidth does not correspond to the value for automatic coupling,
a green bullet is displayed in front of the field.
Compatible Compatible device mode (FSP, FSU, default; default not displayed)