Instrument Functions
R&S
®
FSVA/FSV
251Operating Manual 1176.7510.02 ─ 09
Spectrogram Clear
For spectrogram measurements only.
Resets the Spectrogram result display and clears its history buffer.
Remote command:
CALCulate<n>:SGRam:CLEar[:IMMediate] on page 732
3.2.7.2 Specifying the Sweep Settings
1. Press the Sweep Count softkey and enter the sweep count.
2. Press the Sweeptime Manual or Sweeptime Auto softkey to set the sweep time.
3. Press the Sweep Type softkey to select the sweep type.
4. Press the Sweep Points softkey and enter the number of sweep points.
5. Press the Continuous Sweep or Single Sweep softkey to select the sweep mode.
6. Press the Continue Single Sweep softkey to repeat the single sweep.
3.2.8 Setting Traces – TRACE Key
The TRACE key is used to configure the data acquisition for measurement and the
analysis of the measurement data.
The R&S FSVA/FSV is capable of displaying up to six different traces at a time in a
diagram. A trace consists of a maximum of 691 displayed measurement points on the
horizontal axis (frequency or time). If more measured values than measurement points
are available, several measured values are combined in one displayed measurement
point.
The trace functions include the following:
●
Display mode of the trace
For details on trace modes see Chapter 3.2.8.4, "Trace Mode Overview",
on page 263.
●
Evaluation of the trace as a whole
For details on averaging see Chapter 3.2.8.5, "Description of the Averaging
Method", on page 264.
●
Evaluation of individual measurement points of a trace. For details on detectors
see Chapter 3.2.8.6, "Detector Overview", on page 265.
To open the Trace menu
●
Press the TRACE key.
The "Trace" menu is displayed. The "Trace Configuration" dialog box is displayed.
Menu and softkey description
●
Chapter 3.2.8.1, "Softkeys of the Trace Menu", on page 252
Measurement Parameters