Instrument Functions
R&S
®
FSVA/FSV
235Operating Manual 1176.7510.02 ─ 09
Remote command:
[SENSe:]ADJust:FREQuency on page 777
Auto Level
Defines the optimal reference level for the current measurement automatically.
The measurement time for automatic leveling can be defined using the Settings soft-
key.
You can define a threshold that the signal must exceed before the reference level is
adjusted, see "Upper Level Hysteresis" on page 235 and "Lower Level Hysteresis"
on page 235.
Remote command:
[SENSe:]ADJust:LEVel on page 777
Settings
Opens a submenu to define settings for automatic leveling.
Possible settings are:
"Meas Time Manual" on page 235
"Meas Time Auto" on page 235
Meas Time Manual ← Settings
Opens an edit dialog box to enter the duration of the level measurement in seconds.
The level measurement is used to determine the optimal reference level automatically
(see the "Auto Level" softkey, "Auto Level" on page 235). The default value is 1 ms.
Remote command:
[SENSe:]ADJust:CONFigure:LEVel:DURation on page 778
Meas Time Auto ← Settings
The level measurement is used to determine the optimal reference level automatically
(see the Auto Level softkey).
Upper Level Hysteresis ← Settings
Defines an upper threshold the signal must exceed before the reference level is auto-
matically adjusted when the "Auto Level" function is performed.
Remote command:
[SENSe:]ADJust:CONFiguration:HYSTeresis:UPPer on page 779
Lower Level Hysteresis ← Settings
Defines a lower threshold the signal must exceed before the reference level is auto-
matically adjusted when the "Auto Level" function is performed.
Remote command:
[SENSe:]ADJust:CONFiguration:HYSTeresis:LOWer on page 778
Sweep Type
Opens a submenu to define the sweep type.
This function is not available in IQ Analyzer mode or for input from the R&S Digital I/Q
Interface (option R&S FSV-B17).