Instrument Functions
R&S
®
FSVA/FSV
234Operating Manual 1176.7510.02 ─ 09
Adjusting settings automatically during triggered measurements
When you select an auto adjust function a measurement is performed to determine the
optimal settings. If you select an auto adjust funtion for a triggered measurement, you
can select how the R&S FSVA/FSV should behave:
(default:) The measurement for adjustment waits for the next trigger
The measurement for adjustment is performed without waiting for a trigger.
The trigger source is temporarily set to "Free Run". After the measurement is com-
pleted, the original trigger source is restored. The trigger level is adjusted as fol-
lows:
For IF Power and RF Power triggers:
Trigger Level = Reference Level - 15 dB
For Video trigger:
Trigger Level = 85 %
SCPI command:
[SENSe:]ADJust:CONFigure:TRIG on page 779
Auto All........................................................................................................................234
Auto Freq.................................................................................................................... 234
Auto Level................................................................................................................... 235
Settings....................................................................................................................... 235
Meas Time Manual....................................................................................... 235
Meas Time Auto............................................................................................235
Upper Level Hysteresis.................................................................................235
Lower Level Hysteresis.................................................................................235
Sweep Type................................................................................................................ 235
Sweep...........................................................................................................236
FFT............................................................................................................... 236
Auto...............................................................................................................236
FFT Filter Mode............................................................................................ 236
Auto.................................................................................................... 236
Narrow................................................................................................ 236
Auto All
Performs all automatic settings.
"Auto Freq" on page 234
"Auto Level" on page 235
Remote command:
[SENSe:]ADJust:ALL on page 777
Auto Freq
Defines the center frequency and the reference level automatically by determining the
highest frequency level in the frequency span. This function uses the signal counter;
thus it is intended for use with sinusoidal signals.
This function is not available for input from the R&S Digital I/Q Interface (option
R&S FSV-B17).
This function is not available for 1xEV-DO MS Analysis mode (K85).