Table 4-13: Base unit
Parameter, measuring function or result display Output unit
DBM | DBPW | DBUV | DBMV | DBUA dB (lin/log)
WATT | VOLT | AMPere dB (lin), % (log)
statistics function (APD or CCDF) on dimensionless output
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
CALC:DELT2 ON
Switches on delta marker 2.
CALC:DELT2:Y?
Outputs measurement value of delta marker 2.
Usage: Query only
Manual operation: See "Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
Norm/Delta" on page 280
CALCulate:DELTamarker:SGRam subsystem
The following commands control the deltamarkers when you are using the spectro-
gram.
The spectrogram is available if option R&S FSV-K14 is installed.
CALCulate<n>:DELTamarker<m>:SGRam:FRAMe........................................................... 627
CALCulate<n>:DELTamarker<m>:SGRam:SARea............................................................628
CALCulate<n>:DELTamarker<m>:SGRam:XY:MAXimum[:PEAK].......................................629
CALCulate<n>:DELTamarker<m>:SGRam:XY:MINimum[:PEAK]........................................629
CALCulate<n>:DELTamarker<m>:SGRam:Y:MAXimum:ABOVe........................................ 630
CALCulate<n>:DELTamarker<m>:SGRam:Y:MAXimum:BELow.........................................630
CALCulate<n>:DELTamarker<m>:SGRam:Y:MAXimum:NEXT...........................................631
CALCulate<n>:DELTamarker<m>:SGRam:Y:MAXimum[:PEAK].........................................631
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum:ABOVe..........................................631
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum:BELow.......................................... 632
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum:NEXT............................................632
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum[:PEAK]..........................................633
CALCulate<n>:DELTamarker<m>:SGRam:FRAMe <Frame> | <Time>
This command positions the delta marker on a particular frame. The frame is relative to
the position of marker 1.
The command is available for the spectrogram.