Remote Control
R&S
®
FSVA/FSV
632Operating Manual 1176.7510.02 ─ 09
Suffix:
<n>
.
1...4
Selects the measurement window.
<m> 1...16
Selects the marker.
Example:
CALC:DELT2:SGR:Y:MIN
Activates and positions delta marker 2 on the minimum level of
the y-axis.
CALC:DELT2:SGR:Y:MIN:ABOV
Positions delta marker 2 on the next minimum level found in the
diagram area above the current marker position.
Usage: Event
Manual operation: See "Search Mode for Next Peak in Y Direction" on page 305
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum:BELow
This command positions the specified delta marker on the next minimum level of the
vertical axis. The search includes only frames below the current marker position. It
does not change the horizontal position of the delta marker.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m> 1...16
Selects the marker.
Example:
CALC:DELT2:SGR:Y:MIN
Activates and positions delta marker 2 on the minimum level of
the y-axis.
CALC:DELT2:SGR:Y:MAX:BEL
Positions delta marker 2 on the next minimum level found in the
diagram area below the current marker position.
Usage: Event
Manual operation: See "Search Mode for Next Peak in Y Direction" on page 305
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum:NEXT
This command positions the specified delta marker on the next minimum level of the
vertical axis. The search includes frames above and below the current marker position.
It does not change the horizontal position of the delta marker.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
Remote Control – Commands