Instrument Functions
R&S
®
FSVA/FSV
305Operating Manual 1176.7510.02 ─ 09
"ON"
The minimum frequency included in the peak search range is ≥ 5 ×
resolution bandwidth (RBW).
Due to the interference by the first local oscillator to the first inter-
mediate frequency at the input mixer, the LO is represented as a sig-
nal at 0 Hz. To avoid the peak marker jumping to the LO signal at 0
Hz, this frequency is excluded from the peak search.
"OFF"
No restriction to the search range. The frequency 0 Hz is included in
the marker search functions.
Remote command:
CALCulate<n>:MARKer<m>:LOEXclude on page 666
Search Mode (Spectrograms)
Spectrograms show not only the current sweep results, but also the sweep history.
Thus, when searching for peaks, you must define the search settings within a single
time frame (x-direction) and within several time frames (y-direction).
This setting is only available for spectrogram displays.
Search Mode for Next Peak in X Direction ← Search Mode (Spectrograms)
Selects the search mode for the next peak search within the currently selected frame.
"Left"
Determines the next maximum/minimum to the left of the current
peak.
"Absolute"
Determines the next maximum/minimum to either side of the current
peak.
"Right"
Determines the next maximum/minimum to the right of the current
peak.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum:LEFT on page 668
CALCulate<n>:MARKer<m>:MAXimum:NEXT on page 668
CALCulate<n>:MARKer<m>:MAXimum:RIGHt on page 669
CALCulate<n>:MARKer<m>:MINimum:LEFT on page 670
CALCulate<n>:MARKer<m>:MINimum:NEXT on page 671
CALCulate<n>:MARKer<m>:MINimum:RIGHt on page 671
Search Mode for Next Peak in Y Direction ← Search Mode (Spectrograms)
Selects the search mode for the next peak search within all frames at the current
marker position.
"Up"
Determines the next maximum/minimum above the current peak (in
more recent frames).
"Absolute"
Determines the next maximum/minimum above or below the current
peak (in all frames).
"Down"
Determines the next maximum/minimum below the current peak (in
older frames).
Remote command:
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:ABOVe on page 708
CALCulate<n>:DELTamarker<m>:SGRam:Y:MAXimum:ABOVe on page 630
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:BELow on page 708
Measurement Functions