Remote Control
R&S
®
FSVA/FSV
671Operating Manual 1176.7510.02 ─ 09
Manual operation: See "Next Mode" on page 304
See "Search Mode for Next Peak in X Direction" on page 305
CALCulate<n>:MARKer<m>:MINimum:NEXT
This command positions ae marker to the next higher trace minimum.
If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
In "Spectrogram" mode, this command positions the specified marker on the next mini-
mum level of the x-axis. The search includes the data in both directions of the current
marker position in the current frame. It does not change the position of the marker on
the y-axis.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:MARK2:MIN
Positions marker 2 to the minimum value of the trace.
CALC:MARK2:MIN:NEXT
Positions marker 2 to the next higher maximum value.
Usage: Event
Manual operation: See "Next Min" on page 303
See "Next Mode" on page 304
See "Search Mode for Next Peak in X Direction" on page 305
CALCulate<n>:MARKer<m>:MINimum:RIGHt
This command positions a marker to the next higher trace minimum on the right of the
current value (i.e. in ascending X direction).
If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
In "Spectrogram" mode, this command positions the specified marker on the current
peak level of the x-axis. The search is performed within the current frame. It does not
change the position of the marker on the y-axis.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:MARK2:MIN
Positions marker 2 to the minimum value of the trace.
CALC:MARK2:MIN:RIGH
Positions marker 2 to the next higher minimum value to the right
of the current value.
Usage: Event
Remote Control – Commands