672Operating Manual 1176.7510.02 ─ 09
Manual operation: See "Next Mode" on page 304
See "Search Mode for Next Peak in X Direction" on page 305
CALCulate<n>:MARKer<m>:PEXCursion <Excursion>
This command defines the peak excursion
The peak excursion sets the requirements for a peak to be detected during a peak
search.
The unit depends on the selected operating mode and measurement.
Mode/Display mode Unit
Spectrum dB
ADEMOD, RF display dB
ADEMOD, AM display PCT
ADEMOD, FM display kHz
ADEMOD, PM display RAD
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Parameters:
<Excursion> The peak excursion is the distance to a trace maximum that
must be attained before a new maximum is recognized, or the
distance to a trace minimum that must be attained before a new
minimum is recognized
*RST: 6dB in "Spectrum" mode and RF displays; 5 PCT in
AM displays, 50 kHz in FM displays, (0.5 RAD in
PM displays
Example:
CALC:MARK:PEXC 10dB
Defines peak excursion 10 dB.
Manual operation: See "Peak Excursion" on page 293