Remote Control
R&S
®
FSVA/FSV
593Operating Manual 1176.7510.02 ─ 09
Bit No. Meaning
5 Command Error
This bit is set if a remote command is undefined or has a incorrect syntax.
The R&S FSVA/FSV adds a number to the error queue. The number is in the range between
-100 and -200 and indicates the error type in greater detail.
6 User Request
This bit is set when you press the "Local" softkey.
7 Power On (supply voltage on)
This bit is set when you turn on the instrument.
STATus:OPERation Register
The STATus:OPERation register contains information on current activities of the
R&S FSVA/FSV. It also contains information on activities that have been executed
since the last read out.
You can read out the register with STATus:OPERation:CONDition or STATus:
OPERation[:EVENt].
Table 4-5: Meaning of the bits used in the STATus:OPERation register
Bit No. Meaning
0
CALibrating
This bit is set as long as the instrument is performing a calibration.
1-2 Not used
3
SWEeping
Sweep is being performed; identical to bit 4
4
MEASuring
Measurement is being performed; identical to bit 3
5
Waiting for TRIgger
Instrument is ready to trigger and waiting for trigger signal
6-7 Not used
8 HardCOPy in progress
This bit is set while the instrument is printing a hardcopy.
9 Not used
10
Stop after Range
This bit is set when a range in the sweep list has been completed if "Stop after Range" has
been activated.
11-14 Not used
15 This bit is always 0.
STATus:QUEStionable Register
The STATus:QUEStionable register contains information on instrument states that do
not meet the specifications.
Remote Control - Basics