Remote Control
R&S
®
FSVA/FSV
729Operating Manual 1176.7510.02 ─ 09
Example:
CALC:MATH:MODE LIN
Selects linear averaging for trace math calculations.
Manual operation: See "Lin" on page 259
See "Log" on page 260
See "Power" on page 260
CALCulate:PMETer Subsystem (Power Sensor, R&S FSV-K9)
This subsystem controls the instrument settings for power sensor measurements. It is
only available if the R&S FSV-K9 Power Sensor option is installed.
For details see Chapter 3.9, "Instrument Functions - Power Sensor (R&S FSV-K9)",
on page 538.
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]........................................................... 729
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE.......................................729
CALCulate<n>:PMETer<p>:RELative:STATe....................................................................730
CALCulate<n>:PMETer<p>:RELative[:MAGNitude] <Value>
This command sets the reference value for relative measurements for the selected
power sensor.
Suffix:
<n>
.
irrelevant
<p> 1...4
Power sensor index
Parameters:
<Value> -200 to 200
*RST: 0
Example:
CALC:PMET2:REL -30
Sets the reference value for relative measurements to -30 dBm
for power sensor 2.
Mode: PSM
Manual operation: See "Reference Value" on page 549
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
This command takes the current measurement value as reference value for relative
measurements for the selected power sensor.
Suffix:
<n>
.
irrelevant
<p> 1...4
Power sensor index