Remote Control
R&S
®
FSVA/FSV
730Operating Manual 1176.7510.02 ─ 09
Example:
CALC:PMET2:REL:AUTO ONCE
Takes the current measurement value as reference value for rel-
ative measurements for power sensor 2.
Mode: PSM
Manual operation: See "Meas -> Ref" on page 548
CALCulate<n>:PMETer<p>:RELative:STATe <State>
This command switches between relative and absolute display of the measured power
for the selected power sensor.
Suffix:
<n>
.
irrelevant
<p> 1...4
Power sensor index
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:PMET2:REL:STAT ON
Activates the relative display of the measured value for power
sensor 2.
Mode: PSM
CALCulate:PSEarch|PEAKsearch Subsystem
The CALCulate:PSEarch|PEAKsearch subsystem contains the remote commands for
Spurious Emissions measurements. Both groups of commands (PSEarch and PEAK-
search) perform the same functions.
CALCulate<n>:PEAKsearch|PSEarch[:IMMediate]............................................................ 730
CALCulate<n>:PEAKsearch|PSEarch:AUTO.................................................................... 731
CALCulate<n>:PEAKsearch|PSEarch:MARGin.................................................................731
CALCulate<n>:PEAKsearch|PSEarch:PSHow.................................................................. 731
CALCulate<n>:PEAKsearch|PSEarch:SUBRanges........................................................... 732
CALCulate<n>:PEAKsearch|PSEarch[:IMMediate]
This command switches the spurious limit check off.
If you want to read out the values peak values including the delta to a limit, you have to
switch on the limit again.
This command is only for FSP compatibility, and not necessary to use on the
R&S FSVA/FSV.
Suffix:
<n>
.
irrelevant
Remote Control – Commands