Instrument Functions
R&S
®
FSVA/FSV
433Operating Manual 1176.7510.02 ─ 09
Connections should be as short as possible to keep the usable bandwidth high.
Observe the operating voltage range.
For more information on RT probes, refer to the probe's documentation.
Connecting Active Probes
In order to use active probes with an R&S FSVA/FSV, an RT-ZA9 adapter is required.
The RT-ZA9 provides an interface between the probe's BNC socket and and the ana-
lyzer's N-socket and provides the necessary supply voltages for the probe via the USB
connection. Using this adapter, the following probes are currently supported:
R&S RT-ZS10
RT-ZS10E
RT-ZS20
RT-ZS30
To connect an active probe, proceed as follows:
1. Connect the adapter to the RF Input connector on the R&S FSVA/FSV.
2. Connect the adapter's USB cable to a USB connector on the R&S FSVA/FSV.
3. Connect the probe to the adapter.
Once the probe and adapter have been connected to the R&S FSVA/FSV correctly
and the analyzer has recognized the probe, the "Generic Probe" transducer is activa-
ted and you can start a measurement.
To determine whether the probe has been connected properly and recognized by the
R&S FSVA/FSV, use the remote control command PROB:SET:STAT? (see PROBe:
SETup:STATe? on page 775).
Measurement Functions