EasyManua.ls Logo

Rohde & Schwarz FSVA - Page 775

Rohde & Schwarz FSVA
1106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSVA/FSV
775Operating Manual 1176.7510.02 ─ 09
PROBe:SETup:MODE <Mode>
This command defines which action is taken when the probe's micro button is pressed.
Parameters:
<Mode> RSINgle
A single sweep is performed.
NOACtion
No action is taken.
*RST: OFF
Example:
PROB:SET:STAT ON
Manual operation: See "Probe Config" on page 429
See "Micro Button Action" on page 434
PROBe:SETup:NAME?
This command returns the name of the connected probe.
Example:
PROB:SET:NAME?
Usage: Query only
Manual operation: See "Name" on page 434
PROBe:SETup:STATe?
This command queries whether a probe is connected to the instrument's RF INPUT
and USB connectors and was recognized by the R&S FSVA/FSV.
Example:
PROB:SET:STAT ON
Usage: Query only
PROBe[:STATe] <State>
This command activates a connected probe. Use this command to switch off the probe
and measure the digital input without considering the transducer factor of the probe.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
PROB:STAT ON
Manual operation: See "Probe Config" on page 429
See "State" on page 434