Remote Control
R&S
®
FSVA/FSV
954Operating Manual 1176.7510.02 ─ 09
Parameters:
<Unit> String containing one of the following units:
DB | DBM | DBMV | DBUV | DBUV/M | DBUA DBUA/M | DBPW |
DBPT
*RST: DB
Example:
CORR:TRAN:SEL 'FACTOR'
CORR:TRAN:UNIT 'DBUV'
Selects a transducer based on the unit dBµV.
Manual operation: See "Edit Unit" on page 175
[SENSe:]CORRection:TRANsducer:VIEW <State>
This command includes or excludes a figure of a transducer factor in the diagram.
Before you can use the command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CORR:TRAN:SEL 'FACTOR'
CORR:TRAN:VIEW ON
Selects and displays a transducer factor.
SENSe:ROSCillator Subsystem
This subsystem controls the reference oscillator. The suffix in SENSe is irrelevant for
the commands of this subsystem.
SOURce:EXTernal<1|2>:ROSCillator: EXTernal:FREQuency.............................................954
[SENSe:]ROSCillator:EXTernal:FREQuency.....................................................................954
[SENSe:]ROSCillator:SOURce........................................................................................ 955
[SENSe:]ROSCillator:SOURce:EAUTo?...........................................................................955
SOURce:EXTernal<1|2>:ROSCillator: EXTernal:FREQuency
[SENSe:]ROSCillator:EXTernal:FREQuency <Frequency>
This command defines the frequency of the external reference oscillator.
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Parameters:
<Frequency> Range: 1 MHz to 20 MHz
Example:
ROSC:EXT:FREQ 13
Sets the frequency to 13 MHz.
SOUR:EXT:ROSC:EXT:FREQ 13
Usage: SCPI confirmed
Remote Control – Commands