Remote Control
R&S
®
FSVA/FSV
785Operating Manual 1176.7510.02 ─ 09
[SENSe:]BANDwidth|BWIDth:VIDeo:TYPE <Mode>
This command selects the position of the video filter in the signal path.
Parameters:
<Mode> LINear
The video filter is applied in front of the logarithmic amplifier.
In linear mode, measurements with a logarithmic level scale
result in flatter falling edges compared to logarithmic mode. The
reason is the conversion of linear power values into logarithmic
level values: if you halve the linear power, the logarithmic level
decreases by 3 dB.
LOGarithmic
The video filter is applied after the logarithmic amplifier.
*RST: LINear
Example:
BAND:VID:TYPE LIN
Video filter ahead of the logarithmic amplifier
SENSe:CORRection Subsystem
This subsystem controls calibration and normalization during operation with the track-
ing generator (only with R&S FSVA/FSV option B9 or B10). Furthermore, configuration
of conversion loss tables for external mixers (option B21) is performed using these
commands.
[SENSe:]CORRection:COLLect[:ACQuire]........................................................................785
[SENSe:]CORRection:CVL:BAND....................................................................................786
[SENSe:]CORRection:CVL:CATalog?.............................................................................. 787
[SENSe:]CORRection:CVL:SELect.................................................................................. 787
[SENSe:]CORRection:CVL:MIXer....................................................................................787
[SENSe:]CORRection:CVL:SNUMber.............................................................................. 787
[SENSe:]CORRection:CVL:HARMonic............................................................................. 788
[SENSe:]CORRection:CVL:PORTs..................................................................................788
[SENSe:]CORRection:CVL:BIAS..................................................................................... 789
[SENSe:]CORRection:CVL:COMMent..............................................................................789
[SENSe:]CORRection:CVL:DATA....................................................................................789
[SENSe:]CORRection:CVL:CLEar................................................................................... 790
[SENSe:]CORRection:EGAin:INPut[:MAGNitude]..............................................................790
[SENSe:]CORRection:METHod....................................................................................... 790
[SENSe:]CORRection:RECall..........................................................................................791
[SENSe:]CORRection[:STATe]........................................................................................791
[SENSe:]CORRection:TRANsducer:GENerator.................................................................792
[SENSe:]CORRection:COLLect[:ACQuire] <MeasType>
This command initiates a reference measurement (calibration). The reference mea-
surement is the basis for the measurement normalization.
Remote Control – Commands