Remote Control
R&S
®
FSVA/FSV
790Operating Manual 1176.7510.02 ─ 09
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:DATA 1MHZ,-30DB,2MHZ,-40DB
Mode: all
Manual operation: See "Position/Value" on page 530
[SENSe:]CORRection:CVL:CLEar
This command deletes the selected conversion loss table. Before this command can
be performed, the conversion loss table must be selected (see [SENSe:
]CORRection:CVL:SELect on page 787).
This command is only available with option B21 (External Mixer) installed.
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:CLE
Usage: Event
Mode: all
Manual operation: See "Delete Table" on page 532
[SENSe:]CORRection:EGAin:INPut[:MAGNitude]
This command makes an external gain known to the analyzer, which will take it into
account during the display of measurement results. With this function the gain of an
antenna or of an external preamplifier can be taken into account for the measurement
values.
The command is available if a tracking generator is installed and has been turned on.
Parameters:
Range: -200 dB to 200 dB
*RST: 0 dB
Example:
CORR:EGA:INP 10DB
Takes 10 dB external gain into account.
[SENSe:]CORRection:METHod
This command selects the type of measurement to be performed with the tracking gen-
erator.
The command is available if a tracking generator is installed and has been turned on.
Remote Control – Commands