Instrument Functions
R&S
®
FSVA/FSV
532Operating Manual 1176.7510.02 ─ 09
Shift y ← New Table ← Conversion Loss Table Setup ← External Mixer Config ←
External Mixer
Shifts all conversion loss values by a specific value. The value can be entered in the
edit dialog box. The conversion loss function in the preview pane is shifted along the y-
axis.
Save ← New Table ← Conversion Loss Table Setup ← External Mixer Config ←
External Mixer
The conversion loss table is stored under the specified name in the
C:\r_s\instr\user\cvl\ directory of the instrument.
Edit Table ← Conversion Loss Table Setup ← External Mixer Config ← External
Mixer
Opens the "Edit Conversion Loss Table" dialog box to edit the selected conversion loss
table. For details on table configuration see "New Table" on page 529.
Remote command:
[SENSe:]CORRection:CVL:SELect on page 787
Delete Table ← Conversion Loss Table Setup ← External Mixer Config ← Exter-
nal Mixer
Deletes the currently selected conversion loss table after you confirm the action.
Remote command:
[SENSe:]CORRection:CVL:CLEar on page 790
Import Table ← Conversion Loss Table Setup ← External Mixer Config ← Exter-
nal Mixer
Imports a stored conversion loss table from any directory to the instrument's
C:\r_s\instr\user\cvl\ directory.
Bias Settings ← External Mixer
Opens a dialog box to define the bias current for each range, which is required to set
the mixer to its optimum operating point. It corresponds to the short-circuit current. The
bias current can range from -10 mA to 10 mA. The actual bias current is lower because
of the forward voltage of the mixer diode(s).
The trace is adapted to the settings immediately so you can check the results. To store
the bias setting in the currently selected conversion loss table, select the Write to <CVL
table name> button.
Remote command:
[SENSe:]MIXer:BIAS:LOW on page 823
[SENSe:]MIXer:BIAS:HIGH on page 823
Write to <CVL table name> ← Bias Settings ← External Mixer
Stores the bias setting in the currently selected "Conversion Loss Table" for the range
(see "Conversion Loss Table Setup" on page 528).
Remote command:
[SENSe:]CORRection:CVL:BIAS on page 789
Instrument Functions – External Mixer (Option R&S