"Table"
Defines the conversion loss via the table selected from the list. Pre-
defined conversion loss tables are often provided with the external
mixer and can be imported to the R&S FSVA/FSV. Alternatively, you
can define your own conversion loss tables. Conversion loss tables
are configured and managed in the Conversion Loss Table Setup tab.
For details on conversion loss tables, see Chapter 3.8.1.3, "Conver-
sion Loss Tables", on page 520. For details on importing tables, see
"Import Table" on page 532.
All tables stored on the instrument in the
C:\r_s\instr\user\cvl\ directory are offered for selection. A
validation check is then performed on the selected table to ensure
that it complies with the settings. In particular, the following is
checked:
the assigned band name
the harmonic order
the mixer type
the table must contain at least one frequency that lies within the
frequency range for the band
Remote command:
[SENSe:]MIXer:LOSS[:LOW] on page 828
Average for range 1
[SENSe:]MIXer:LOSS:TABLe[:LOW] on page 828
Table for range 1
[SENSe:]MIXer:LOSS:HIGH on page 827
Average for range 2
[SENSe:]MIXer:LOSS:TABLe:HIGH on page 827
Table for range 2
Conversion Loss Table Setup ← External Mixer Config ← External Mixer
In this tab you configure and manage conversion loss tables. Conversion loss tables
consist of value pairs that describe the correction values for conversion loss at certain
frequencies. The correction values for frequencies between the reference points are
obtained via interpolation.
The currently selected table for each range is displayed at the top of the dialog box.