Remote Control
R&S
®
FSVA/FSV
827Operating Manual 1176.7510.02 ─ 09
[SENSe:]MIXer:HARMonic:TYPE <OddEven>
This command specifies whether the harmonic order to be used should be odd, even,
or both.
Parameters:
<OddEven> ODD | EVEN | EODD
*RST: EVEN
Example:
MIX:HARM:TYPE ODD
Mode: all
Manual operation: See "Harmonic Type" on page 527
[SENSe:]MIXer:LOPower
<Level>
This command specifies the LO level of the external mixer's LO port.
Parameters:
<Level> <numeric value>
Range: 13.0 dBm to 17.0 dBm
Increment: 0.1 dB
*RST: 15.5 dBm
Example:
MIX:LOP 16.0dBm
Mode: all
Manual operation: See "LO Level" on page 533
[SENSe:]MIXer:LOSS:HIGH <Average>
This command defines the average conversion loss to be used for the entire high
range.
Parameters:
<Average> <numeric value>
Range: 0 to 100
*RST: 40.0 dB
Default unit: dB
Example:
MIX:LOSS:HIGH 20dB
Mode: all
Manual operation: See "Conversion Loss" on page 527
[SENSe:]MIXer:LOSS:TABLe:HIGH <FileName>
This command defines the file name of the conversion loss table to be used for the
high range.