Remote Control
R&S
®
FSVA/FSV
826Operating Manual 1176.7510.02 ─ 09
Usage: Event
Mode: all
Manual operation: See "Preset Band" on page 526
[SENSe:]MIXer:HARMonic:HIGH:STATe <State>
This command specifies whether a second (high) harmonic is to be used to cover the
band's frequency range.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
MIX:HARM:HIGH:STAT ON
Mode: all
Manual operation: See "Range 1/2" on page 527
[SENSe:]MIXer:HARMonic:HIGH[:VALue] <HarmOrder>
This command specifies the harmonic order to be used for the high range.
Parameters:
<HarmOrder> <numeric value>
Range: 2 to 100 (USER band); for other bands: see band
definition
Example:
MIX:HARM:HIGH 2
Mode: all
Manual operation: See "Harmonic Order" on page 527
[SENSe:]MIXer:HARMonic[:LOW] <HarmOrder>
This command specifies the harmonic order to be used for the LOW range.
Parameters:
<HarmOrder> <numeric value>
Range: 2 to 100 (USER band); for other bands: see band
definition
*RST: 2 (for band F)
Example:
MIX:HARM 3
Mode: all
Manual operation: See "Harmonic Order" on page 527
Remote Control – Commands