Remote Control
R&S
®
FSVA/FSV
825Operating Manual 1176.7510.02 ─ 09
Usage: Query only
Mode: all
Manual operation: See "RF Start" on page 526
[SENSe:]MIXer:FREQuency:STOP?
This command queries the frequency at which the external mixer band stops.
Example:
MIX:FREQ:STOP?
Queries the stop frequency of the band.
Usage: Query only
Mode: all
Manual operation: See "RF Stop" on page 526
[SENSe:]MIXer:HARMonic:BAND <Band>
This command selects the external mixer band. The query returns the currently
selected band.
This command is only available if the external mixer is active (see [SENSe:]MIXer[:
STATe] on page 823).
Parameters:
<Band> A | KA | Q | U | V | E | W | F | D | G | Y | J | USER
Standard waveguide band or user-defined band.
Note: The band formerly referred to as "A" is now named "KA";
the input parameter "A" is still available and refers to the same
band as "KA".
For a definition of the frequency range for the pre-defined bands,
see Table 3-22).
*RST: F (90 GHz - 140 GHz)
Example:
MIX ON
Activates the external mixer.
MIX:HARM:BAND A
Sets the band to A (26.5 GHz - 40 GHz).
Mode: all
Manual operation: See "Band" on page 526
[SENSe:]MIXer:HARMonic:BAND:PRESet
This command restores the preset frequency ranges for the standard waveguide
bands.
Example:
MIX:HARM:BAND:PRES
Presets the waveguide bands.
Remote Control – Commands