Remote Control
R&S
®
FSVA/FSV
886Operating Manual 1176.7510.02 ─ 09
SYSTem:COMMunicate:TCPip:RDEVice:GENerator<generator>:ADDRess
<Address>
Configures the TCP/IP address for the external tracking generator 1 or 2.
This command is only available if the R&S FSVA/FSV option B10 (External Tracking
Generator) is installed.
Suffix:
<generator>
.
1 | 2
tracking generator
For Noise Figure measurements (K30): 1 only
Parameters:
<Address> TCP/IP address between 0.0.0.0 and 0.255.255.255
*RST: 0.0.0.0
Example:
SYST:COMM:TCP:RDEV:GEN1:ADDR 130.094.122.195
Mode: A
SYSTem:SPEaker:VOLume <Volume>
This command defines the volume of the built-in loudspeaker for demodulated signals.
The command is available with the audio demodulator (option R&S FSV-B3) and in the
time domain in Spectrum mode.
Parameters:
<Volume> Range: 0 to 1
*RST: 0.5
Example:
SYST:SPE:VOL 0
Switches the loudspeaker to mute.
SYSTem:TIME:TSTamp?
This command queries the time and date from the internal real time clock.
Return values:
<TimeStamp>
Example:
SYST:TIME:TSTamp?
Usage: Query only
4.2.3.19 TRACe Subsystem
The TRACe subsystem controls access to the instruments internal trace memory.
Some applications offer up to four measurement windows. For those, the suffix <n> for
TRACe selects the measurement window. For all other applications, the suffix is irrele-
vant.
Remote Control – Commands