EasyManua.ls Logo

Rohde & Schwarz FSVA - Page 828

Rohde & Schwarz FSVA
1106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSVA/FSV
828Operating Manual 1176.7510.02 ─ 09
Parameters:
<FileName> <file name>
Example:
MIX:LOSS:HIGH 20dB
Mode: all
Manual operation: See "Conversion Loss" on page 527
[SENSe:]MIXer:LOSS:TABLe[:LOW] <FileName>
This command defines the file name of the conversion loss table to be used for the low
range.
Parameters:
<FileName> <file name>
Example:
MIX:LOSS:TABL mix_1_4
Specifies the conversion loss table mix_1_4.
Mode: all
Manual operation: See "Conversion Loss" on page 527
[SENSe:]MIXer:LOSS[:LOW] <Average>
This command defines the average conversion loss to be used for the entire low range.
Parameters:
<Average> <numeric value>
Range: 0 to 100
*RST: 40.0 dB
Default unit: dB
Example:
MIX:LOSS 20dB
Mode: all
Manual operation: See "Conversion Loss" on page 527
[SENSe:]MIXer:PORTs <PortType>
This command specifies whether the mixer is a 2-port or 3-port type.
Parameters:
<PortType> 2 | 3
*RST: 2
Example:
MIX:PORT 3
Mode: all
Manual operation: See "Mixer Type" on page 527
Remote Control – Commands

Table of Contents