Remote Control
R&S
®
FSVA/FSV
828Operating Manual 1176.7510.02 ─ 09
Parameters:
<FileName> <file name>
Example:
MIX:LOSS:HIGH 20dB
Mode: all
Manual operation: See "Conversion Loss" on page 527
[SENSe:]MIXer:LOSS:TABLe[:LOW] <FileName>
This command defines the file name of the conversion loss table to be used for the low
range.
Parameters:
<FileName> <file name>
Example:
MIX:LOSS:TABL mix_1_4
Specifies the conversion loss table mix_1_4.
Mode: all
Manual operation: See "Conversion Loss" on page 527
[SENSe:]MIXer:LOSS[:LOW] <Average>
This command defines the average conversion loss to be used for the entire low range.
Parameters:
<Average> <numeric value>
Range: 0 to 100
*RST: 40.0 dB
Default unit: dB
Example:
MIX:LOSS 20dB
Mode: all
Manual operation: See "Conversion Loss" on page 527
[SENSe:]MIXer:PORTs <PortType>
This command specifies whether the mixer is a 2-port or 3-port type.
Parameters:
<PortType> 2 | 3
*RST: 2
Example:
MIX:PORT 3
Mode: all
Manual operation: See "Mixer Type" on page 527
Remote Control – Commands