Instrument Functions
R&S
®
FSVA/FSV
530Operating Manual 1176.7510.02 ─ 09
Remote command:
[SENSe:]CORRection:CVL:BAND on page 786
Harmonic Order ← New Table ← Conversion Loss Table Setup ← External Mixer
Config ← External Mixer
The harmonic order for which the table is to be applied.
Remote command:
[SENSe:]CORRection:CVL:HARMonic on page 788
Bias ← New Table ← Conversion Loss Table Setup ← External Mixer Config ←
External Mixer
The bias current which is required to set the mixer to its optimum operating point. It
corresponds to the short-circuit current. The bias current can range from -10 mA to 10
mA. The actual bias current is lower because of the forward voltage of the mixer
diode(s).
Tip: You can also define the bias interactively while a preview of the trace with the
changed setting is displayed, see "Bias Settings" on page 532.
Remote command:
[SENSe:]CORRection:CVL:BIAS on page 789
Mixer Name ← New Table ← Conversion Loss Table Setup ← External Mixer
Config ← External Mixer
Specifies the name of the external mixer for which the table is to be applied.
Remote command:
[SENSe:]CORRection:CVL:MIXer on page 787
Mixer S/N ← New Table ← Conversion Loss Table Setup ← External Mixer Config
← External Mixer
Specifies the serial number of the external mixer for which the table is to be applied.
Remote command:
[SENSe:]CORRection:CVL:SNUMber on page 787
Mixer Type ← New Table ← Conversion Loss Table Setup ← External Mixer Con-
fig ← External Mixer
Specifies whether the external mixer for which the table is to be applied is a two-port or
three-port type.
Remote command:
[SENSe:]CORRection:CVL:PORTs on page 788
Position/Value ← New Table ← Conversion Loss Table Setup ← External Mixer
Config ← External Mixer
Each position/value pair defines the correction value for conversion loss for a specific
frequency. The reference values must be entered in order of increasing frequencies. A
maximum of 50 reference values can be entered.
Instrument Functions – External Mixer (Option R&S
FSV-B21)