Use Zoom Limits ← Search Limits
Restricts the marker search to the zoomed area.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits:ZOOM on page 675
Search Lim Off ← Search Limits
Deactivates all limits of the search range.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe] on page 674
CALCulate<n>:THReshold:STATe on page 742
Next Mode
Selects the mode of the Next Peak or Next Min softkey.
Three settings are available:
"<"
Sets the active marker/delta marker to the next maximum/minimum
left to the marker of the selected trace.
"abs"
Sets the active marker/delta marker to the next lower maximum/
higher minimum of the selected trace.
">"
Sets the active marker/delta marker to the next maximum/minimum
right to the marker of the selected trace.
Remote command:
Next Peak:
CALC:MARK:MAX:LEFT (<): CALCulate<n>:MARKer<m>:MAXimum:LEFT
on page 668
CALCulate<n>:DELTamarker<m>:MAXimum:LEFT on page 622
CALC:MARK:MAX:RIGH (>): CALCulate<n>:MARKer<m>:MAXimum:RIGHt
on page 669
CALCulate<n>:DELTamarker<m>:MAXimum:RIGHt on page 622
CALC:DELT:MAX:NEXT (abs): CALCulate<n>:MARKer<m>:MAXimum:NEXT
on page 668
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT on page 622
Next Min:
CALC:MARK:MIN:LEFT (>): CALCulate<n>:MARKer<m>:MINimum:LEFT
on page 670
CALCulate<n>:DELTamarker<m>:MINimum:LEFT on page 623
CALC:MARK:MIN:RIGH (>): CALCulate<n>:MARKer<m>:MINimum:RIGHt
on page 671
CALCulate<n>:DELTamarker<m>:MINimum:RIGHt on page 624
CALC:MARK:MIN:NEXT (abs): CALCulate<n>:MARKer<m>:MINimum:NEXT
on page 671
CALCulate<n>:DELTamarker<m>:MINimum:NEXT on page 624
Exclude LO
Switches the frequency range limit for the marker search functions on or off.