Remote Control
R&S
®
FSVA/FSV
622Operating Manual 1176.7510.02 ─ 09
CALCulate<n>:DELTamarker<m>:MAXimum:LEFT
This command positions the delta marker to the next smaller trace maximum on the left
of the current value (i.e. descending X values). The corresponding delta marker is acti-
vated first, if necessary.
If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
In "Spectrogram" mode, this command positions the specified delta marker on the next
peak level of the x-axis. The search includes only the data to the left of the current
marker position in the current frame. It does not change the position of the deltamarker
on the y-axis.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:DELT:MAX:LEFT
Sets delta marker 1 to the next smaller maximum value to the
left of the current value.
Manual operation: See "Next Mode" on page 304
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT
This command positions the delta marker to the next smaller trace maximum. The cor-
responding delta marker is activated first, if necessary.
If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
In "Spectrogram" mode, this command positions the specified delta marker on the next
peak level of the x-axis. The search includes the data in both directions of the current
marker position in the current frame. It does not change the position of the delta
marker on the y-axis.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:DELT2:MAX:NEXT
Sets delta marker 2 to the next smaller maximum value.
Manual operation: See "Next Peak" on page 302
See "Next Mode" on page 304
CALCulate<n>:DELTamarker<m>:MAXimum:RIGHt
This command positions the delta marker to the next smaller trace maximum on the
right of the current value (i.e. ascending X values). The corresponding delta marker is
activated first, if necessary.