Instrument Functions
R&S
®
FSVA/FSV
302Operating Manual 1176.7510.02 ─ 09
└ Search Lim Off..............................................................................................304
Next Mode...................................................................................................................304
Exclude LO..................................................................................................................304
Search Mode (Spectrograms).....................................................................................305
└ Search Mode for Next Peak in X Direction................................................... 305
└ Search Mode for Next Peak in Y Direction................................................... 305
└ Marker Search Type..................................................................................... 306
└ Marker Search Area......................................................................................306
Peak Excursion........................................................................................................... 306
Select Marker (No)
Opens a submenu to select one of 16 markers and define whether the marker is a nor-
mal or a delta marker (see "Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
Norm/Delta" on page 280). "(No)" indicates the number of the currently active marker.
Peak
Sets the active marker/delta marker to the highest maximum of the trace.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum[:PEAK] on page 667
CALCulate<n>:DELTamarker<m>:MAXimum[:PEAK] on page 621
Next Peak
Sets the active marker/delta marker to the next maximum of the selected trace.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum:NEXT on page 668
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT on page 622
Center =Mkr Freq (span > 0)
Sets the center frequency to the current marker or delta marker frequency. A signal
can thus be set to as center frequency, for example to examine it in detail with a
smaller span.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:CENTer on page 680
Ref Lvl =Mkr Lvl
Sets the reference level to the current marker level.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence on page 692
Marker to Trace
Opens an edit dialog box to enter the number of the trace on which the marker is to be
placed.
Remote command:
CALCulate<n>:MARKer<m>:TRACe on page 673
CALCulate<n>:DELTamarker<m>:TRACe on page 625
Min
Sets the active marker/delta marker to the minimum of the selected trace.
Measurement Functions