Remote Control
R&S
®
FSVA/FSV
692Operating Manual 1176.7510.02 ─ 09
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:FUNC:NOIS ON
Switches on the noise measurement.
Manual operation: See "Noise Meas On/Off" on page 288
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult
This command queries the result of the noise measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK2 ON
Switches on marker 2.
CALC:MARK:FUNC:NOIS ON
Switches on noise measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:NOIS:RES?
Outputs the noise result of marker 2.
Manual operation: See "Noise Meas On/Off" on page 288
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
This command sets the reference level to the power measured by a marker.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:MARK2:FUNC:REF
Sets the reference level to the level of marker 2.
Manual operation: See "Ref Lvl =Mkr Lvl" on page 302