Loading...
Instrument Functions
R&S
®
FSVA/FSV
288Operating Manual 1176.7510.02 ─ 09
n dB down................................................................................................................... 292
Marker Peak List......................................................................................................... 293
Peak List On/Off............................................................................................293
Sort Mode Freq/Lvl....................................................................................... 293
Max Peak Count........................................................................................... 293
Peak Excursion.............................................................................................293
Left Limit....................................................................................................... 293
Right Limit.....................................................................................................294
Threshold......................................................................................................294
ASCII File Export.......................................................................................... 294
Decim Sep.................................................................................................... 294
Marker Number.............................................................................................294
Band Power.................................................................................................................295
Select Marker (No)........................................................................................295
Band Power On/Off.......................................................................................295
Span..............................................................................................................295
Power............................................................................................................295
Density..........................................................................................................295
Select Marker (No)
Opens a submenu to select one of 16 markers and define whether the marker is a nor-
mal or a delta marker (see "Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
Norm/Delta" on page 280). "(No)" indicates the number of the currently active marker.
See "Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker Norm/Delta" on page 280.
Signal Count
Switches the frequency counter on or off, and opens an edit dialog box to define the
resolution of the frequency counter, if enabled. The frequency is counted at the posi-
tion of the reference marker (marker 1). If no marker is activate, marker 1 is switched
on and positioned on the largest signal.
The sweep stops at the reference marker until the frequency counter has delivered a
result. The result is displayed in the marker field (see Figure 3-9), labeled with [Tx
CNT].
For more information see Chapter 3.3.3.5, "Frequency Measurement with the Fre-
quency Counter", on page 297.
Remote command:
CALCulate<n>:MARKer<m>:COUNt on page 664
CALCulate<n>:MARKer<m>:COUNt:FREQuency? on page 665
Noise Meas On/Off
Switches the noise measurement for the active marker on or off. The corresponding
marker becomes the normal marker.
For more information on noise measurement see Chapter 3.3.3.6, "Measurement of
Noise Density", on page 297.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe] on page 691
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult on page 692