Remote Control
R&S
®
FSVA/FSV
665Operating Manual 1176.7510.02 ─ 09
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Parameters:
<State> ON | OFF
*RST: OFF
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 1.
CALC:MARK:COUN ON
Switches on the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value.
Manual operation: See "Signal Count" on page 288
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
This command queries the result of the frequency counter for marker 1.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Return values:
<Frequency> Frequency at the marker position.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 2.
CALC:MARK:COUN ON
Activates the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value of marker 1.
Usage: Query only
Manual operation: See "Signal Count" on page 288
Remote Control – Commands