Remote Control
R&S
®
FSVA/FSV
664Operating Manual 1176.7510.02 ─ 09
CALCulate<n>:MARKer<m>[:STATe] <State>
This command turns markers on and off.
If the corresponding marker number is currently active as a deltamarker, it is turned
into a normal marker.
Suffix:
<n>
.
Selects the measurement window.
<m> depends on mode
Selects the marker.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK3 ON
Switches on marker 3 or switches to marker mode.
Manual operation: See "Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
Norm/Delta" on page 280
See "Marker Wizard" on page 281
CALCulate<n>:MARKer<m>:AOFF
This command all markers off, including delta markers and marker measurement func-
tions.
Suffix:
<n>
.
Selects the measurement window.
<m> depends on mode
irrelevant
Example:
CALC:MARK:AOFF
Switches off all markers.
Usage: Event
Manual operation: See "All Marker Off" on page 282
CALCulate<n>:MARKer<m>:COUNt <State>
This command turns the frequency counter at the marker position for marker 1 on and
off.
The count result is queried with CALCulate<n>:MARKer<m>:COUNt:FREQuency?.
The frequency counter works for one marker only. If you perform a frequency count
with another marker, the R&S FSVA/FSV deactivates the frequency count of the first
marker.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Remote Control – Commands

Table of Contents