Instrument Functions
R&S
®
FSVA/FSV
295Operating Manual 1176.7510.02 ─ 09
Band Power
Opens a submenu to activate and configure a band power marker. Band power mark-
ers allow you to measure the integrated power for a defined span (band) around a
marker. The result can be displayed either as a power (dBm) or density (dBm/Hz). The
span is indicated by lines in the diagram.
Band power markers are only available for standard frequency measurements in Spec-
trum mode.
Note: As of firmware version 2.0, Band Power markers are also available in the I/Q
Analyzer, but only for Spectrum displays (see "Display Config" on page 447).
For more information see Chapter 3.3.3.8, "Performing Band Power Measurements",
on page 299.
Select Marker (No) ← Band Power
Opens a submenu to select one of 16 markers and define whether the marker is a nor-
mal or a delta marker (see "Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
Norm/Delta" on page 280). "(No)" indicates the number of the currently active marker.
See "Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker Norm/Delta" on page 280.
Band Power On/Off ← Band Power
Activates or deactivates the band power marker. When switched to on, if no marker is
active yet, marker 1 is activated. Otherwise, the currently active marker is used as a
band power marker (all other marker functions for this marker are deactivated). All
markers can be defined as band power markers, each with a different span.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer[:STATe] on page 679
Span ← Band Power
Defines the span (band) around the marker for which the power is measured. The span
is indicated by lines in the diagram.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:SPAN on page 679
Power ← Band Power
Selects the power mode for the band power marker, i.e. the result is displayed in dBm.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:MODE on page 678
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:RESult? on page 678
Density ← Band Power
Selects the density mode for the band power marker, i.e. the result is displayed in
dBm/Hz.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:MODE on page 678
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:RESult? on page 678
Measurement Functions