Instrument Functions
R&S
®
FSVA/FSV
306Operating Manual 1176.7510.02 ─ 09
CALCulate<n>:DELTamarker<m>:SGRam:Y:MAXimum:BELow on page 630
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:NEXT on page 709
CALCulate<n>:DELTamarker<m>:SGRam:Y:MAXimum:NEXT on page 631
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:ABOVe on page 709
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum:ABOVe on page 631
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:BELow on page 710
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum:BELow on page 632
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:NEXT on page 710
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum:NEXT on page 632
Marker Search Type ← Search Mode (Spectrograms)
Defines the type of search to be performed in the spectrogram.
"X-Search"
Searches only within the currently selected frame.
"Y-Search"
Searches within all frames but only at the current marker position.
"XY-Search"
Searches in all frames at all positions.
Remote command:
CALCulate<n>:MARKer<m>:SGRam:XY:MAXimum[:PEAK] on page 707
CALCulate<n>:DELTamarker<m>:SGRam:XY:MAXimum[:PEAK] on page 629
CALCulate<n>:MARKer<m>:SGRam:XY:MINimum[:PEAK] on page 707
CALCulate<n>:DELTamarker<m>:SGRam:XY:MINimum[:PEAK] on page 629
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum[:PEAK] on page 709
CALCulate<n>:DELTamarker<m>:SGRam:Y:MAXimum[:PEAK] on page 631
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum[:PEAK] on page 711
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum[:PEAK] on page 633
CALCulate<n>:MARKer<m>:MAXimum[:PEAK] on page 667
CALCulate<n>:DELTamarker<m>:MAXimum[:PEAK] on page 621
CALCulate<n>:MARKer<m>:MINimum[:PEAK] on page 669
CALCulate<n>:DELTamarker<m>:MINimum[:PEAK] on page 623
Marker Search Area ← Search Mode (Spectrograms)
Defines which frames the search is performed in.
"Visible"
Only the visible frames are searched.
"Memory"
All frames stored in the memory are searched.
Remote command:
CALCulate<n>:MARKer<m>:SGRam:SARea on page 706
CALCulate<n>:DELTamarker<m>:SGRam:SARea on page 628
Peak Excursion
Opens an edit dialog box for level measurements to enter the minimum level value by
which a signal must rise or fall so that it will be identified as a maximum or a minimum
by the search functions. Entries from 0 dB to 80 dB are allowed; the resolution is 0.1
dB. The default setting for the peak excursion is 6 dB.
For more information see "Specifying the suitable peak excursion" and "Effect of differ-
ent peak excursion settings".
Remote command:
CALCulate<n>:MARKer<m>:PEXCursion on page 672