Remote Control
R&S
®
FSVA/FSV
706Operating Manual 1176.7510.02 ─ 09
CALCulate<n>:MARKer<m>:SGRam:FRAMe................................................................... 706
CALCulate<n>:MARKer<m>:SGRam:SARea....................................................................706
CALCulate<n>:MARKer<m>:SGRam:XY:MAXimum[:PEAK]...............................................707
CALCulate<n>:MARKer<m>:SGRam:XY:MINimum[:PEAK]................................................707
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:ABOVe................................................ 708
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:BELow.................................................708
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:NEXT...................................................709
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum[:PEAK].................................................709
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:ABOVe................................................. 709
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:BELow..................................................710
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:NEXT....................................................710
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum[:PEAK]..................................................711
CALCulate<n>:MARKer<m>:SGRam:FRAMe <Frame> | <Time>
This command positions the marker on a particular frame.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m> 1...16
Selects the marker.
Parameters:
<Frame> | <Time> <Frame> (if time stamp is off)
Defines the frame to place the marker on. The range is {0...num-
ber of recorded frames-1}
<time> (if time stamp is on)
Defines the distance of the marker in seconds to the last mea-
sured frame (frame 0).
Example:
CALC:MARK:SGR:FRAM -20
Sets the marker on the 20th frame before the present.
CALC:MARK2:SGR:FRAM 2 s
Sets second marker on the frame 2 seconds ago.
CALCulate<n>:MARKer<m>:SGRam:SARea <SearchArea>
This command defines the marker search area.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m> 1...16
Selects the marker.
Remote Control – Commands