Remote Control
R&S
®
FSVA/FSV
710Operating Manual 1176.7510.02 ─ 09
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m> 1...16
Selects the marker.
Example:
CALC:MARK2:SGR:Y:MIN
Activates and positions marker 2 on the minimum level of the y-
axis.
CALC:MARK2:SGR:Y:MIN:ABOV
Positions marker 2 on the next minimum level found in the dia-
gram area above the current marker position.
Usage: Event
Manual operation: See "Search Mode for Next Peak in Y Direction" on page 305
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:BELow
This command positions the specified marker on the next minimum level of the vertical
axis. The search includes only frames below the current marker position. It does not
change the horizontal position of the marker.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m> 1...16
Selects the marker.
Example:
CALC:MARK2:SGR:Y:MIN
Activates and positions marker 2 on the minimum level of the y-
axis.
CALC:MARK2:SGR:Y:MAX:BEL
Positions marker 2 on the next minimum level found in the dia-
gram area below the current marker position.
Usage: Event
Manual operation: See "Search Mode for Next Peak in Y Direction" on page 305
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:NEXT
This command positions the specified marker on the next minimum level of the vertical
axis. The search includes frames above and below the current marker position. It does
not change the horizontal position of the marker.
The command is available for the spectrogram.